A laboratory is accredited to measure dissolved lead in drinking water by inductively coupled plasma mass spectrometry (ICP-MS). Can it issue an accredited result for dissolved lead in wastewater using the same instrument? The answer depends on whether the published scope of accreditation covers that matrix, the method used and any limitations that apply.

A scope of accreditation describes the specific activities for which an accreditation body has recognised a laboratory’s competence. For a testing laboratory accredited to ISO/IEC 17025, the scope helps customers identify which activities accreditation covers. It also helps laboratory staff decide which samples to accept, which methods to use and which results they can report as accredited.

Whether you are preparing an initial application or extending an existing scope, start with the tests your customers need and define each activity at a level your laboratory can support with evidence.

1. Start with the tests your customers need

Review your test register, recent customer requests, contracts and any regulatory programmes that require accredited results. For each proposed activity, identify the types of samples customers submit, the results they need and any prescribed methods or reporting limits that affect whether those results will be accepted.

For example, a customer may need a lead result at a concentration below your current reporting limit. That matters when deciding what performance the laboratory needs to demonstrate before including the test in its application.

Prioritise activities for which there is a clear need and supporting evidence is available. If an important test still needs verification or validation, decide whether to complete that work before applying or include the test in a later scope extension.

2. Describe each activity precisely

Use your accreditation body’s application template and terminology. Prepare a working list of the proposed activities so you can check the wording against your methods and technical records. The details shown on the published scope may vary by accreditation body and testing discipline.

  • Material or matrix. Identify the sample type, such as drinking water, wastewater or soil. Use a broader grouping only where the supporting evidence covers the proposed grouping and the accreditation body accepts the wording.
  • Analyte, property or test. Specify what the laboratory determines, such as dissolved lead, pH or tensile strength. Include distinctions that affect the meaning of the result.
  • Method and technique. Identify the standard method or controlled in-house procedure and, where relevant, the analytical technique. Include sample preparation details where they define what is measured.
  • Range and limitations. Identify the supported measurement range, reporting limit or other restriction where relevant to the activity or required by the accreditation body.
  • Location and sampling. Identify where the activity is performed. If you want sampling or field testing covered, describe those activities explicitly in the application.

Worked example — Dissolved lead in drinking water

Suppose a laboratory has demonstrated performance for dissolved lead in drinking water from 1 to 100 µg/L using an in-house ICP-MS procedure. Its working entry could read as follows.

Material: Drinking water
Analyte: Dissolved lead
Technique and method: ICP-MS, in-house procedure MET-07
Supported range: 1–100 µg/L
Location: Main laboratory

The wording “metals in water by ICP-MS” leaves staff and customers guessing which metals, water types and concentration ranges are covered. The detailed entry lets staff see immediately that they need to check for separate wastewater coverage before accepting a wastewater sample as accredited work.

The range and procedure identifier in this example are illustrative. Use your laboratory’s own evidence and your accreditation body’s format when drafting an entry.

3. Match every proposed entry to technical evidence

Create an internal register with one row per proposed activity. For each row, record the method identifier, the matrices and analytes covered, and where to find the supporting records. An assessor or technical reviewer should be able to trace the proposed wording back to the evidence.

Review the following records as applicable.

  • Method verification or validation. Show that the laboratory can achieve the required performance for the proposed sample types, analytes, range and intended use.
  • Personnel, equipment and facilities. Identify authorised staff and the equipment, facilities and locations that support the activity.
  • Metrological traceability and measurement uncertainty. Identify the calibration or reference material records that establish traceability and the relevant measurement uncertainty evaluation.
  • Ongoing validity of results. Identify routine quality controls and relevant proficiency testing or other interlaboratory comparisons.

Pay particular attention to sample preparation. Measuring dissolved metals in filtered water and measuring metals after digesting an unfiltered sample can produce results with different meanings. Check that your proposed scope, method records and report wording describe the same measurement.

If the proposed wording goes beyond the evidence, narrow the entry or define the additional work needed to support it. For example, adding wastewater may require investigating matrix effects and demonstrating that the method still meets the required performance criteria.

If you need to collect more evidence before extending the scope, see our guide on how to develop a method validation plan.

4. Decide how changes to the scope will be managed

A fixed scope typically identifies specific accredited activities. A flexible scope, where offered and granted by the accreditation body, may allow certain changes within agreed boundaries under the laboratory’s controlled process. The scope granted and the accreditation body’s rules determine how method revisions, additional analytes, new matrices and new locations are handled.

Suppose a drinking-water client asks you to add another metal to an existing ICP-MS panel. Before offering the new result as accredited, check whether the current scope permits the addition, what evidence is needed and whether the accreditation body must approve a scope extension.

Keep the published scope and your internal test register aligned with quotation and reporting systems. When a change takes effect, staff should be able to tell which version applies and which work they can report as accredited.

5. Test the wording against a real customer request

Give the draft scope and a representative enquiry to a colleague who handles quotations, sample receipt or report review. Ask them to answer three questions.

  1. Does the scope cover the sample type and the result the customer has requested?
  2. Can the laboratory meet the requested reporting limit using a covered method and location?
  3. If the order includes both covered and uncovered activities, can staff identify which results may be reported as accredited?

If the answer depends on asking the method developer what an entry means, review the wording or document how the scope maps to individual tests in your internal systems. Run the same check after a scope extension, especially when the change affects how work is quoted, assigned or reported.

A practical scope gives customers a clear description of the accredited services they can request and gives staff a reliable basis for deciding what they can accept and report. Before submitting an application, check that every proposed entry serves both purposes and is supported by technical evidence.

For more on how technical competence and ongoing controls fit into a laboratory quality system, read our guide to laboratory quality management.

Preparing an initial application or adding tests to an existing scope? Explore ISO4LAB’s laboratory consulting and technical support services.